Fracture resistance of distal extension cantilevers of implant supported prostheses

Objectives: To determine the fracture resistance to linear vertical compressive forces of the cantilever arms of acrylic-only and metal-reinforced acrylic fixed implant supported prosthesis.

Methods: Ten non-reinforced and 10 reinforced acrylic superstructures were secured on five evenly distributed implants and subjected to linear axial compressive forces utilising an Instron® materials testing machine. The reinforcement consisted of commercially available preformed titanium metal strengthener bars. Force was applied to the cantilever arms 15mm distal to the distal most implant, and two tests were conducted on each sample. The first drop in load recorded was noted as a fracture of the sample.

Results: Fracture of the acrylic was noted at the distal most implant in both non-reinforced and the reinforced samples. The mean fracture value for the non-reinforced samples was 679N and for the reinforced samples, 628N.

Conclusion: No significant difference between the fracture resistance of the two designs was noted.

Supported in part by Southern Implants (Irene, South Africa), and the Faculty of Health Sciences, University of the Witwatersrand.


Division: South African Division
Meeting: 2006 South African Division (Midrand, South Africa)
Location: Midrand, South Africa
Year: 2006
Final Presentation ID:
Abstract Category|Abstract Category(s): Scientific Groups
Authors
  • Osswald, Martin A.  ( University of the Witwatersrand, Johannesburg, N/A, South Africa )
  • Owen, C. Peter  ( University of the Witwatersrand, Johannesburg, N/A, South Africa )
  • Howes, Dale G.  ( University of the Witwatersrand, Johannesburg, N/A, South Africa )
  • SESSION INFORMATION
    Oral Session
    Prosthodontics