Delivered Irradiance as a function of LCU-Tip to Restoration Surface

Objectives: To quantify the delivered irradiance as a function of distance of the light curing unit (LCU) tip from the restoration surface and different LCUs. Methods: Six LCUs, bluephase-20i, bluephase-G2, bluephase-style (Ivoclar-Vivadent); freelight2, EliparS10 (3MESPE); VALO (Ultradent), were individually mounted on a MARC - Resin Calibrator™ and centred perpendicularly over the sensor. The distance between the LCU tip and the sensor was increased from 0.00mm initially in 1.00mm increments up to a distance of 10.00mm. At each distance, the LCU was activated for 10 seconds and the energy per unit area (mW/cm2) delivered to the sensor recorded. Results: Table1. Irradiance of all six LCUs decreased as a function of distance. bluephase G2 had the least variance over distance (0.00 to 10.00mm). 3M ESPE LCUs showed a wide divergent beam with a 73-82% loss in irradiance over 10mm, similar to the bluephase20i with turbo tip. Conclusions: Reporting a single irradiance value or a single energy spectral range to describe the output from a curing light is both imprecise and inappropriate. LCU manufacturers should indicate irradiance values delivered at different clinically relevant T-S distances. Information of both the irradiance distribution, and the distribution of the spectral emission across the light tip should be provided. Recommendations for curing times for different clinical T-S distances should be given by manufacturers of RBCs.

Table 1. The Change in Irradiance* with T-S distance (0 – 10mm) and the overall % Changes in Irradiance from 0.00mm to 10.00mm.

bluephase G2

bluephase 20i (turbo tip)

bluephase style

freelight2

Elipar S10

VALO standard

VALO

high

VALO plasma

manufacturer irradiance

1200

(+/- 5%)

2000

(+/- 5%)

1100

(+/- 5%)

1000

(+/- 5%)

1200

(+/- 5%)

1000

(+/- 5%)

1400

(+/- 5%)

3200

(+/- 5%)

irradiance at 0 mm

1200.3[19.7]

1855.8[33.6]

1255.5[23.0

1676.5[55.4]

2141.8[20.7]

1343.5[29.9]

1974.5[41.9]

3479.5[58.3]

irradiance at 1 mm

1227.8[28.7]

1723.8[63.2]

1283.3[39.2]

1572.8[22.5]

2041.5[38.1]

1388.2[48.8]

2080.5[20.9]

3515.5[91.3]

Irradiance change from 0-1.00 (%)

+2.3%

-7.1%

+2.2%

-6.2%

-4.7%

+3.3%

+5.4%

+1%

irradiance at 2 mm

1287.0[49.9

1533.3[50.6]

1365.1[66.1]

1391.1[30.2]

1886.2[40.8]

1466.3[37.1]

2120.3[52.3]

3516.3[193.0]

Irradiance change from 0-2.00 (%)

+7.2%

-17.4%

+8.7%

-17%

-11.9%

+9.1%

+7.4%

+1%

irradiance at 3 mm

1347.2[38.8]

1303.3[40.9]

1408.5[48.6]

1167.0[66.9]

1682.0[36.9

1372.2[44.9]

1980.0[28.1]

3280.8[128.8

Irradiance change from 0-3.00 (%)

+12.2%

-29.8%

+12.2%

-30.4%

-21.5%

+2.1%

+

30.3%

-5.7%

irradiance at 4 mm

1392.5[30.3]

1087.2[53.3]

1362.5[37.9]

943.5[58.3]

1430.0[47.6]

1163.3[33.8]

1696.8[41.3]

2928.3[207.3]

Irradiance change from 0-4.00 (%)

+16%

-41.3%

+8.5%

-43.7%

-33.2%

-13.4%

-14.1%

-15.8%

irradiance at 6 mm

1395.8[45.3]

723.0[33.7]

1066.0[73.6]

614.3[42.2]

992.5[32.8]

848.5[18.1]

1250.2[15.1]

2200.0[184.6]

Irradiance change from 0-6.00 (%)

+16.3%

-61%

-15.1%

-63.4%

-53.7%

-36.8%

-36.7%

-36.8%

irradiance at 8 mm

1225.2[34.1]

466.0[22.6]

754.5[55.6]

422.00[24.3]

721.8[30.8]

677.0[20.6]

1000.3[17.1]

1764.8[145.6]

Irradiance change from 0-8.00 (%)

+2.1%

-74.9%

-39.9%

-74.8%

-66.3%

-49.6%

-49.3%

-49.3%

irradiance at 10 mm

976.5[26.3]

319.0[10.9]

542.0[43.5]

300.3[17.00]

565.8[23.3]

567.5[13.2

836.3[9.8]

1468.8[106.6]

Irradiance change from 0-10 (%)

-18.6%

-82.8%

-56.8%

-82.1%

-73.6%

-57.8%

-57.6%

-57.8%


Division: Pan European Region Meeting
Meeting: 2014 Pan European Region Meeting (Dubrovnik, Croatia)
Location: Dubrovnik, Croatia
Year: 2014
Final Presentation ID: 580
Abstract Category|Abstract Category(s): Scientific Groups
Authors
  • Almuallem, Zahra  ( THE UNIVERSITY OF EDINBURGH, EDINBURGH, , Scotland )
  • Santini, Ario  ( THE UNIVERSITY OF EDINBURGH, EDINBURGH, , Scotland )
  • SESSION INFORMATION
    Oral Session
    Light-curing and Endodontology
    09/13/2014