IADR Abstract Archives

Agreement of working length in C-shaped canal permanent molar determined by an electronic apex locator at 0.5 bar in model for root length determination and actual working length

Objectives: The aim of this study was to examine the agreement of a working length in C-shaped canal permanent molar determined by an electronic apex locator at 0.5 bars in model and actual working length.
Methods: Extracted human permanent molar teeth of c-shaped canal group (42 teeth) and non c-shaped canal group (42 teeth) were included in the study. All teeth were measured working length by Root ZX® apex locator in model then actual working length was done under stereomicroscope (30X) after clearing process. The data were statistically analyzed by Intraclass Correlation Coefficient (ICC) which number is close to 1 means in a good reliability.
Results: The result showed that ICC in c-shaped canal group and non c-shaped canal group were in a good reliability (ICC = 0.948 and 0.940). ICC in c-shaped canal groups classified by Fan’s classification were 0.944, 0.928 and 0.949 in C1, C2 and C3 respectively. Within C-shaped canal group, both c-shaped canal and straight canal showed a good reliability (ICC= 0.947, 0.939).
Conclusions: In conclusion working length determined by an electronic apex locator and actual working length in clearing process in all groups were in a good correlation.

2017 Japanese Division Annual Meeting (Tokyo, Japan)

2017

Dental Materials 6: Instruments and Equipment
  • Jainaen, Angsana  ( Faculty of Dentistry, Khon Kaen University , Muang , Khon Kaen , Thailand )
  • None
    Poster Session
    Abstracts Presented