Anodized vs. Sandblasted Implant Surfaces in Grafted-Dehiscence Defect Animal Model
Objectives: The effect of impact of surface characteristics on osseointegration and bone apposition were investigated by comparing a novel gradient anodised (NGA) functionalized, commercially available, implant (NobelActive™, TiUltra NP, 3.5 × 8.5 mm, Nobel Biocare AG, Switzerland) with custom-made replicas of the NGA implant geometry, modified with the SLActive and SLA surface (Replica of NobelActive, SLActive™ or SLA™, Roxolid, 3.5 x 8.5 mm, Institut Straumann AG, Switzerland). Methods: Crestal bone formation was analyzed histologically and histomorphometrically using a standardized 3mm acute type buccal dehiscence model grafted with deproteinized bovine bone mineral (DBBM-Bio-Oss™,Geistlich®) and covered with a porcine collagen membrane (Bio-Gide™,Geistlich®) after 2 (2w,n=6) and 8weeks(8w,n=8) of healing. Each animal received one NGA, SLActive and SLA implant. The primary outcome was bone-to-implant contact in the dehiscence defect(dBIC). Secondary outcomes included: first-bone-to-implant-contact(fBIC), new-bone-height(NBH), vertical-bone-creep(VBC) and bone-area-to-total-area(BATA). A multivariate linear regression model, with the factor “animal” modelled as a random effect, was adjusted for multiple comparison using Dunnett-Hsu method with a significance level of α<0.05. Results: At 2w, dBIC was statistically significantly higher for SLActive(µ=17.1,p<0.05) compared to NGA(µ=2.5). At 8w both SLActive(µ=47.8,p<0.001) and SLA(µ=35.1, p<0.01) demonstrated significantly higher dBIC than NGA(µ=3.2). At both time points, fBIC was significant closer to the crest for SLActive(2w,µ=-14072,p<0.01) (8w,µ=-461.3,p<0.01) and SLA(2w,µ=-1971.5, p<0.05) (8w,µ=-652.1,p<0.01) compared to NGA(2w,µ=-2567.1)(8w,µ=-1633.5). There was significantly more VBC for SLActive(2w,µ=735.6,p<0.05)(8w,µ=1728.2,p<0.01) and SLA(2w,µ=509.0,p<0.05)(8w,µ=1757.9,p<0.01) at 2w and 8w compared to NGA(2w,µ=-69.0)(8w,µ=629.1). At 2w and 8w there was no evidence of a difference(p>0.05) in NBH or BATA between the groups. Conclusions: At 2 and 8 weeks the SLActive surface demonstrated significantly higher levels of dBIC, fBIC and VBC compared to NGA. SLA surface demonstrated significantly higher levels of fBIC and VBC at 2 weeks and dBIC, fBIC and VBC at 8 weeks compared to the NGA surface.
2023 IADR/LAR General Session with WCPD 2023 0371 Implantology
Gill, Thomas
( Queen Mary Univeristy of London
, London
, United Kingdom
; Barts Health NHS Trust
, London
, United Kingdom
)
Bosshardt, Dieter
( University of Bern
, Bern
, Switzerland
)
Roccuzzo, Mario
( University of Torino
, Torino
, Italy
)
Shahdad, Shakeel
( Queen Mary Univeristy of London
, London
, United Kingdom
; Barts Health NHS Trust
, London
, United Kingdom
)
The present study was funded by a grant from Institut Straumann AG, implants and test implant were provided by Straumann.
TG declares no conflict of interest.
DB declares no conflict of interest.
MR and SS receive speaker honorariums for educational cour