Objective: The purpose of this study is to see if the biaxial flexural strength test of a light-cured composite can be used to validate the EOP (Energy for Optimum Polymerization) data obtained previously via Differential Scanning Calorimetry (DSC). Methods: Renew® hybrid composite (Bisco), VIP visible light curing unit (Bisco) and an Instron Model 4465 were used. Test specimens of 13mm diameter and 2mm thickness were light polymerized with a 13mm light probe at 1mm distance with VIP at 500mW/cm2. The composite was placed in an SS split mold and covered with Mylar at both sides and cured. All specimens were stored in water @ 37ºC for 24 or 72 hrs prior to testing. [Each specimen was placed on top of three steel balls forming a triangle on the biaxial assembly base. A special biaxial shield was placed around the specimen and head to align the head properly.] Results: (n=20) (Units in MPa) Analysis of the data by one way ANOVA followed by a multiple comparison using a Student-Newman-Keuls test.
Cure time |
5 sec. |
7 sec. |
10 sec. |
15 sec. |
20 sec. |
40 sec. |
80 sec. |
Storage time |
|
|
|
|
|
|
|
24 hr. |
71.2 |
101.5 |
117.8 |
125.4 |
114.4 |
131.3 |
122.1 |
|
(14.5) |
(14.3) |
(25.3) |
(30.0) |
(23.7) |
(22.7) |
(25.7) |
72 hr. |
71.2 |
89.8 |
116.5 |
113.2 |
128.7 |
128.8 |
122.9 |
|
(16.5) |
(7.9) |
(20.8) |
(26.3) |
(27.1) |
(21.3) |
(23.8) |
Conclusions: For both 24 and 72 hr. storage, biaxial flexural strength reaches a constant after about 10sec of irradiation. Biaxial flexural strength can be used as an indication of cure for EOP determinations.