Impact of CBCT-Artefacts and Exposure Settings on Implant Measurement Accuracy

Objectives: To evaluate the effect of metallic artefact and kVp levels on the measurement accuracy of implants and osteotomy sites in two CBCT machines in-vitro.
Methods: A radiographically visible block was prepared. Five implants with progressive thread design (diameter: 3.5mm, length: 11mm) were placed perpendicular to the block. A groove was carved at 2mm apical to the implants to represent the mandibular canal. The block was scanned using two CBCT machines: Machine A (Kodak-9000, 70kVp, 7mA, 0.2mm voxel-size) and Machine B (I-CAT, 120kVp, 5mA, 0.3mm voxel-size). The exposure factors were set at the recommended factory presets. The implants were then removed; the osteotomy length was measured using a digital caliper at the center of the site and scanned in an identical manner. The scans were analyzed separately using iCATVision and CS3D-imaging software, with each software’s default window width and level settings. Measurements of the implant and osteotomy sites were done by two blinded examiners; mean values were recorded and subjected to statistical analysis using linear mixed model at 5% significance.
Results: No statistically significant difference for implant measurements was observed between the two machines independent of kVp-levels. However, significant difference was found in the osteotomy measurements of scans from Machine A, acquired at lower kVp-level than machine B (P< 0.0001), due to better visualization of remaining bone at the apical part of the osteotomy during implant removal (Table).
Conclusions: CBCT exposure parameters likely impact the effect of implant-caused metallic artefacts on the accuracy of measurements at surgical implant sites.
Division: IADR/AADR/CADR General Session
Meeting: 2019 IADR/AADR/CADR General Session (Vancouver, BC, Canada)
Location: Vancouver, BC, Canada
Year: 2019
Final Presentation ID: 0781
Abstract Category|Abstract Category(s): Implantology Research
Authors
  • Mahdian, Mina  ( Stony Brook University , Stony Brook , New York , United States )
  • Patchell, Gayle  ( Stony Brook University , Stony Brook , New York , United States )
  • Colosi, Dan  ( Stony Brook University , Stony Brook , New York , United States )
  • Hou, Wei  ( Stony Brook University , Stony Brook , New York , United States )
  • Nam, Kyung  ( Stony Brook University , Stony Brook , New York , United States )
  • Romanos, Georgios  ( Stony Brook University , Stony Brook , New York , United States )
  • Financial Interest Disclosure: None
    SESSION INFORMATION
    Poster Session
    Emerging Technologies in Implantology
    Thursday, 06/20/2019 , 11:00AM - 12:15PM
    TABLES
     Machine-A(mm)Machine-B(mm)P-Value
     Mean ± SD
    Median (Min, Max)
    Mean ± SD
    Median (Min, Max)
     
    Implant Length0.2±0.2
    0.2 (0.0,0.6)
    0.2±0.1
    0.2 (0.1,0.4)
    0.78
    Implant Diameter0.4±0.3
    0.3 (0.1,0.9)
    0.4±0.2
    0.4 (0.0,0.5)
    0.9
    Distance to Canal0.2±0.2
    0.2 (0.0,0.6)
    0.3±0.2
    0.3 (0.1,0.6)
    0.52
    Osteotomy1.3±0.6
    1.4 (0.2,2.1)
    0.7±0.4
    0.7 (0.0,1.7)
    <0.0001